Neuromodulation Cuff Production Protocol
Observed a cutting-edge neuromodulation cuff fabrication procedure and developed a rigorous protocol to document it in detail.

Problem / Goal
huMannity was developing a nerve stimulation cuff for a new indication that required a significantly smaller, more intricate design than prior iterations. The fabrication process was complex, often spanning multiple days per batch, and existed entirely in one engineer’s head with no written documentation. Without a formal protocol, the process could not be transferred, scaled, or validated.

What I Did
I shadowed the fabrication engineer through multiple production runs, documenting each step with detailed notes and researching the underlying materials science to assign precise technical language to procedures that had never been formally named.
Making the cuffs myself gave me the hands-on understanding needed to interrogate the process. The work included casting and molding medical-grade silicones, dispensing adhesives with an EFD dispenser, hand-forming 0.005” Pt-Ir electrode wire under a microscope, and laser welding cables to electrodes embedded in the cuff body. I systematically varied key parameters across curing time and temperature, EFD tip selection, and laser welder settings to replace judgment calls with defined, reproducible values.
Results
The final protocol ran 55 pages and covered each fabrication step with quantitative parameters, annotated photographs, and Keyence microscope imagery. It was written to document control standards and is intended to serve as a training reference for technicians producing the cuff.
What I Learned
Starting with no background in materials science or silicone fabrication, I had to build enough domain knowledge to make the cuffs myself before writing about them credibly. The work sharpened my sense of how to turn expert intuition into transferable documentation.